Purchasers Include:

< Click

NITROGEN ANALYZER SPECIFICATIONS:

A digital electronic meter to measure
nitrogen gas (volume %) in respiratory air.
Includes sampling valve and vacuum pump.
For research only. Not for patient diagnosis.

  • Cabinet dimensions: 12" wide X 10" deep X 4½" high.
  • Color: gray with black front.
  • Power cord: 8 foot length.
  • Electrical requirements: 120 volt, 500 milliamp fuse.
  • Transducer: gas ionization, light filter and detector.
  • Range: 0 to 80% nitrogen gas (in oxygen).
  • Digital readout: to 0.1% nitrogen.
  • Recorder output: 1 Volt (DC) per 10% nitrogen.
  • Accuracy: ± 1% within range (software corrected).
  • Linearity: graph is provided with each instrument.
  • Response: 95% response in < 50 ms.
  • Warm-up time: approx. 30 minutes.
  • Warranty: one year on parts and labor.